Multipurpose & Endoscopic Ultrasound Phantom ATS 570

The Model 570 Multipurpose & Endoscopic phantom is an easy, comprehensive means of evaluating imaging systems with an operating frequency range of 2.25 to 7.5 MHz. The phantom is designed with a combination of monofilament line targets for distance measurements and tissue mimicking target structures of varying sizes and contrasts. Due to the acoustic similarity of the background material and the target structures, artifacts caused by distortion, shadowing and enhancement have been eliminated. Four gray scale targets ranging in contrast from +6 to -3 dB are provided to evaluate the system’s displayed dynamic range and gray scale processing performance.

The Model 570 offers a new and improved scan surface design to easily accommodate linear, sector, endoscopic probes and mechanical sector probes such as used for rectal scanning.

Scan Surface

Multipurpose & Endoscopic Ultrasound Phantom ATS 570 front view Multipurpose & Endoscopic Ultrasound Phantom ATS 570 top view Multipurpose & Endoscopic Ultrasound Phantom ATS 570 rear view
Standard Scan Surface Endoscopic Scan Surface Scanning Well

Tests Performed:
  • Dead Zone or Ring-Down (A1 & A2)
  • Horizontal Measurement Calibration Linear (C1 & C2)
  • Horizontal Measurement Calibration Sector (C1 & C2)
  • Sensitivity/Penetration (F)
  • Vertical Measurement Calibration (B1 & B2)
  • Image Uniformity
  • Functional Resolution (F1, F2)
  • Focal Zone (B1, B2, or F)
  • Axial & Lateral Resolution (E1, E2, E3)
  • Contrast, Gray Scale (G1, G2, G3)
Uses:
  • Quality Assurance
  • Routine Clinical Performance
  • In-process Manufacturing
  • Final Q.A. Inspection
  • Research & Development
  • Teaching
Specifications
General
Overall Dimensions 270 x 215 x 96 cm
Housing Material PVC
Scan Surfaces 2
Scan Wells 1
Weight 4.55 Kg (10 lb)
Tissue Mimicking Material
Type Urethane Rubber
Freezing Point < -40°C
Melting Point > 100°C
Attenuation Coefficient 0.5 dB/cm/MHz ± 5.0%
Speed of Sound 1450 m/s ± 1.0% at 23°
Line Targets
Material
Diameter
Position Tolerance
Monofilament Nylon
.12 mm
± 0.1 mm
Dead Zone Groups
Lateral Displacement
Interval Spacing
Scan Surface Depth
Standard
Endoscopic
A1
5 mm
1 mm
 
2-10 mm
A2
5 mm
1 mm

 
2-10 mm
Vertical Groups
Number of Targets
Interval Spacing
Depth
B1
17
10 mm
10-160 mm
B2
8
10 mm
10-80 mm
Horizontal Linear
Groups
Number of Targets
Interval Spacing
Scan Surface Depth
Standard
Endoscopic
C1
 
5 mm
10 mm
 
40 mm
C2
 
5 mm
110 mm
 
 
40 mm
Axial-Lateral Resolution Groups
 
 
Number of Targets
 
Interval Spacing
 
Scan Surface Depth
E1
Standard
E2
Endoscopic
E3
Scan Well
Horizontal Linear
Groups
Number of Targets
Interval Spacing
Scan Surface Depth
Standard
Endoscopic
D1
 
5
10 mm
  
50 mm
D2
 
5
10 mm
  

50 mm
6
5, 4, 3, 2, 1 mm
40 mm 60 mm 40 mm
Anechoic Target Structures F1 F2 Gray Scale Target Structures G1
Standard
G2
Endoscopic
G3
Scan Well
Type Non-echogenic, cylindrical Type Echogenic, Cylindrical
Size 8, 4, 2 mm 6 mm Diameters 10 mm 10 mm 6 mm
Number of Targets 27 1 Scan Surface Depth 30 mm 20 mm 30 mm
Depth 10 - 170 mm 30 mm Number of Targets 4 4 2
Interval Spacing 10 & 20 mm 30 mm Contrast relative to background material (dB) +6, +3, -3, -6 +3, -3

Multipurpose & Endoscopic Ultrasound Phantom ATS 570 Scan Surface Diagram


Buy It Now