Multipurpose Ultrasound Phantom ATS 539

The Model 539 Multipurpose phantom is an easy, comprehensive means of evaluating imaging systems with an operating frequency range of 2.25 to 7.5 MHz.

The phantom is designed with a combination of monofilament line targets for distance measurements and tissue mimicking target structures of varying sizes and contrasts. Cystic-like target structures are positioned in-line vertically, thereby permitting an entire target group to be displayed in one view. Due to the acoustic similarity of the background material and the target structures, artifacts caused by distortion, shadowing or enhancement have been eliminated.

Six gray scale targets ranging in contrast from +15 to -15 dB are provided to evaluate the system’s displayed dynamic range and gray scale processing performance.

Tests Performed
  • Dead Zone or Ring-Down
  • Vertical Measurement Calibration
  • Horizontal Measurement Calibration
  • Sensitivity/Penetration
  • Functional Resolution
  • Focal Zone
  • Axial & Lateral Resolution
  • Image Uniformity
  • Gray Scale & Displayed Dynamic Range
Uses
  • Quality Assurance
  • Routine Clinical Performance
  • In-process Manufacturing
  • Final Q.A. Inspection
  • Research & Development
  • Teaching

Specifications

General

Overall Dimensions 23.4 x 20.5 x 9.5 cm*
Weight 3.1 Kg*
Housing Material PVC
Wall Thickness 1.0 cm*
Scan Surfaces 4
Scan Surface
Dimensions
17.5 x 7.5 cm*
14.0 x 7.5 cm*

Tissue Mimicking Material

Type Urethane Rubber
Freezing Point <-40°C
Melting Point > 100°C
Attenuation Coefficient 0.5 dB/cm/MHz ± 5.0%**
Speed of Sound 1450 m/s ± 1.0% at 23°

Line Targets

Material Monofilament Nylon
Diameter 0.12mm ± 5%
Position Tolerance ± 0.1 mm
Vertical Group  
Number of Targets 17
Interval Spacing 1.0 cm
Depth 1.0 - 18.0 cm
Horizontal Group  
Number of Targets 10
Number of Groups 2
Interval Spacing 2.0 cm
Depth
Scanning Surface #1
Scanning Surface #3
 
5.0 cm
5.0 cm

 

Dead Zone Group  
Number of Targets 9
Lateral Displacement 5.0 mm
Interval Spacing 1.0 mm
Depth 2.0 - 10.0 mm
Axial-Lateral Resolution Group  
Number of Targets 11
Lateral Displacement 1.0 mm
Interval Spacing 5.0, 4.0, 3.0, 2.0, 1.0 mm
Depth 7.0, 11.0, 4.0, 16.0 cm

Anechoic Target Structures

Type Non-echogenic, cylindrical
Positional Tolerance ± 0.1 mm
Diameter Tolerance ± 5%
Number of Groups 5
Targets:
Group 2, 3, 4
Group 6,8
 
17
8
Diameters
Group 2, 3, 4
Group 6, 8
 
1.0 cm
2.0 cm
Depth 1.0 - 17.0 cm

Gray Scale Target Structures

Type Echogenic, Cylindrical
Number of Targets 6
Diameters 15.0 mm
Interval Spacing (center to center) 2.0 cm
Depth 4.0 cm
Contrast relative to background material (dB) +15, +6, +3, -3, -6, -15
*Nominal Dimensions
**Other attenuations are available upon request
**Lifetime Warranty**
Carrying Case & Built-in Scan Wells Included

Multipurpose Ultrasound Phantom ATS 539 Target Diagram


Buy It Now