Resolution – Penetration Ultrasound Phantom ATS 534

The Model 534 tissue-mimicking phantom™ (see also Model 504) is designed to evaluate the performance of diagnostic ultrasound imaging systems. These phantoms contain 81 anechoic, cylindrical target structures of varying sizes. The targets are positioned in-line vertically, to permit an entire target group to be displayed in one view. Due to the acoustic similarity of the background material and the anechoic targets, artifacts caused by distortion, shadowing or enhancement have been eliminated.

The Model 534 utilizes a rubber-based™ material with a Lifetime warranty.

Uses
  • Quality Assurance
  • Basic Clinical Performance
  • In-process Manufacturing
  • Teaching
Tests Performed
  • Sensitivity
  • Focal Zone
  • Functional Resolution
  • Image Uniformity

SPECIFICATIONS: Model 534
General
Overall Dimensions: 20.3 x 23.3 x 9.5 cm
Scan Surfaces: 4
Scan Surface Dimensions: 14.5 x 7.5 cm
Scan Surface Dimensions: 17.5 x 7.5 cm
Weight: 3.1 Kg
Tissue Mimicking Material
Type: Urethane Rubber
Freezing Point: <-40°C
Melting Point: >100°C
Speed of Sound: 1450 m/s ± 1.0% @ 23°C
Attenuation Coefficient: 0.5 dB/cm/MHz ± 5.0%
Anechoic Target Structures
Type: Non-echogenic, cylindrical
Diameters: 2, 3, 4, 6, 8, 10 mm
Depth: 1 - 18 CM
Includes Carrying Case & Built-in Scan Wells.
Resolution – Penetration Ultrasound Phantom ATS 534 Target Diagram